AFM/SPM PROBE
(원자현미경/주사탐침현미경 탐침)
Product Description | |||
---|---|---|---|
The new PointProbe®Plus (PPP) combines the well-known features of the proven PointProbe® series such as high application versatility and compatibility with most commercial SPMs with a more reproducible tip radius as well as a more defined tip shape. The minimized variation in tip shape provides more reproducible images.
The PPP-EFM probe is offered for electrostatic force microscopy. An overall metallic coating (PtIr5) on both sides of the cantilever increasing the electrical conductivity of the tip. The force constant of this type is specially tailored for the electrostatic force microscopy yielding very high force sensitivity while simultaneously enabling tapping mode and lift mode operation.
The probe offers unique features: metallic conductivity of the tip high mechanical Q-factor for high sensitivity precise alignment of the cantilever position (within +/- 2 µm) when used with the Alignment Chip compatible with PointProbe® Plus XY-Alignment Series
Please note: Wear at the tip can occur if operating in contact-, friction- or force modulation mode. Although this is possible, it is not recommended to use PtIr5 coated tips for electrical contacting in applications where it is necessary to conduct high current. The very thin layer of PtIr5 is unable to support much current.
It is compatible with: Bruker / Veeco / Digital Instruments Keysight / Agilent / Molecular Imaging Asylum Research Park Systems JEOL JPK etc. |
|||
Coating Description | |||
The PtIr5 coating is an approximately 25 nm thick double layer of chromium and platinum iridium5 on both sides of the cantilever. The tip side coating enhances the conductivity of the tip and allows electrical contacts. The detector side coating enhances the reflectivity of the laser beam by a factor of about 2 and prevents light from interfering within the cantilever. The coating process is optimized for stress compensation and wear resistance. The bending of the cantilever due to stress is less than 3.5% of the cantilever length. | |||
AFM Tip | |||
Shape | Standard | ||
Height | 10-15um | ||
Setback | 10-15um | ||
Radius | <25nm | ||
AFM Cantilever | |||
Shape | Beam | ||
Length | 225um (215-235um) | ||
Width | 28um (20-35um) | ||
Thickness | 3um (2-4um) | ||
Force Constant | 2/m (0.5-9.5N/m) | ||
Resonance Frequency | 75kHz (45-115kHz) | ||